Product Preset Therapies
TL1000 Interferential Stimulator

Std (Standard Mode) - The unit provides a beat frequency of 1 Hz for 6 sec. and then supplies a beat frequency of 10 Hz for 6 sec. This therapy is repeated for 10 min. After the initial 10 min. treatment the unit produces a beat frequency that sweeps from 80 Hz to 150 Hz within 6 sec. This second treatment continues for another 10 min. After 20 min., the unit shuts off. ( For pain and swelling after surgery)

Swp (Sweep) - The unit provides a beat frequency that sweeps from 1 Hz to 150 Hz within 6 sec. This treatment continues for 30 min. After 30 min. the unit shuts off. (For Rehab)

Back (Back) - The unit provides a beat frequency of 1 Hz for 6 sec. and then a beat frequency of 10 Hz for 6 sec. This therapy is repeated for 60 min. after which the unit shuts off. (For back injuries)

Cont (Continuous) - The unit provides a beat frequency of 100 Hz for the full treatment time of 60 min.

Stim (Stimulator) - The unit provides a beat frequency of 48 hz for 6 sec. followed by a ramp down to "0" amplitude that lasts for 6 sec. This phase is followed by a ramp back to the original amplitude where the cycle is repeated. This treatment lasts for 60 min.

Mem1, Mem2, Mem3, Mem4 (Memory Locations) - The unit allows modifications of the five previously described therapies. These customized therapies can be stored into four memory locations.

TL1100 Interferential Stimulator

Std (Standard Mode) - The unit provides a beat frequency of 1 Hz for 6 sec. and then supplies a beat frequency of 10 Hz for 6 sec. This therapy is repeated for 10 min. After the initial 10 min. treatment the unit produces a beat frequency that sweeps from 80 Hz to 150 Hz within 6 sec. This second treatment continues for another 10 min. After 20 min., the unit shuts off. ( For pain and swelling after surgery)

36R (30% - 60% Ramp ) - The unit provides a beat frequency that gradually changes (ramps) from 30% below the set frequency to 60% above the set frequency within 6 sec. This treatment continues for 30 min. After 30 min. the unit shuts off.

36A (30% - 60% Abrupt) - The unit provides a beat frequency that abruptly changes from 30% below the set frequency to 60% above the set frequency within 6 sec. This treatment continues for 30 min. After 30 min. the unit shuts off.

Cont (Continuous) - The unit provides a beat frequency of 100 Hz for the full treatment time of 60 min.

Stim (Stimulator) - The unit provides a beat frequency of 48 hz for 6 sec. followed by a ramp down to "0" amplitude that lasts for 6 sec. This phase is followed by a ramp back to the original amplitude where the cycle is repeated. This treatment lasts for 60 min.

Mem (Memory Locations) - The unit allows modifications of one of the five previously described therapies. The customized therapies can be stored into a memory locations and recalled by the patient for use.

TL2000 Galvanic Stimulator

DAY (Day Mode) - The unit provides 100 PPS continuous for 15 min., then 200 PPS continuous for another 15 min. After 30 min. the unit shuts off. Depending in the polarity of the output leads, the user will receive treatment for chronic and acute pain.

NGT (Night Mode) - The unit provides 100 PPS continuous negative polarity for 15 min., then 20 PPS continuous negative polarity for another 15 min. The unit is then idle for 30 min. for a treatment cycle time of 60 min. The unit then repeats this cycle for five more times for a total treatment time of six hours.

FSP (Full Sweep) - The unit sweeps from 1 to 150 PPS continuous negative polarity over six seconds. It then sweeps back from 150 to 1 PPS continuous negative polarity over a second six second period. It continues providing this therapy for 30 min After 30 min. the unit shuts off.

MSP (Muscle Pump) - The unit provides a smooth amplitude ramp of 50 PPS continuous positive polarity from zero amplitude to the amplitude setting provided by the user. This ramp takes place in six seconds. Then, the unit provides a smooth amplitude ramp from the amplitude setting provided by the user to zero in a one half second period. The unit is then remains at zero amplitude for one second. This cycle is then repeated for a total treatment time of 20 Min.

M1- M3 (Memory Locations 1 through 3) - The unit allows modification of one of the four previously described therapies. This customized therapy can be stored into any of three memory location and recalled by the patient for use.