TL2000 Preset Therapies

DAY (Day Mode) - The unit provides 100 PPS continuous for 15 min., then 200 PPS continuous for another 15 min. After 30 min. the unit shuts off. Depending in the polarity of the output leads, the user will receive treatment for chronic and acute pain.

NGT (Night Mode) - The unit provides 100 PPS continuous negative polarity for 15 min., then 20 PPS continuous negative polarity for another 15 min. The unit is then idle for 30 min. for a treatment cycle time of 60 min. The unit then repeats this cycle for five more times for a total treatment time of six hours.

FSP (Full Sweep) - The unit sweeps from 1 to 150 PPS continuous negative polarity over six seconds. It then sweeps back from 150 to 1 PPS continuous negative polarity over a second six second period. It continues providing this therapy for 30 min After 30 min. the unit shuts off.

MSP (Muscle Pump) - The unit provides a smooth amplitude ramp of 50 PPS continuous positive polarity from zero amplitude to the amplitude setting provided by the user. This ramp takes place in six seconds. Then, the unit provides a smooth amplitude ramp from the amplitude setting provided by the user to zero in a one half second period. The unit is then remains at zero amplitude for one second. This cycle is then repeated for a total treatment time of 20 Min.

M1- M3 (Memory Locations 1 through 3) - The unit allows modification of one of the four previously described therapies. This customized therapy can be stored into any of three memory location and recalled by the patient for use.